- 3 -
Steam Sterilization (Autoclaving)
PENTAX water bottle assemblies, Model OS-H4 have been designed to withstand high-pressure steam sterilization
procedures providing the following parameters are adhered to:
During steam sterilization, ensure that the cap and tubing section have been removed from the water bottle container.
Make sure that the drain lever on the water bottle cap has been set on the A/W position (upright).
Sterilizer Type: Prevacuum
Temperature: 132 ~ 135 °C (270 ~ 275 °F)
Time: 5 minutes