Permolock C3 Owner´s Manual Troubleshooting
4040
The light on the locking
plate control button flashes
red.
EVENT POSSIBLE CAUSE ACTION
Troubleshooting guide
The following troubleshooting guide describes a number of errors and events
which could occur when using your Permolock C3, together with suggested
remedies. Please note that this guide does not describe all the problems and
incidents that may arise. You should always contact your service contact or
Permobil in the event of uncertainty.
Electronics fault Reset system by pressing the
control button for 10 seconds.
Contact service
The light on the locking
plate control button shines
red during manual release.
The locking mechanism is
not in the locked position.
Make sure the locking mecha-
nism is in locked position and
that the wheelchair is fully in or
out of the docking system.
The docking system does
not unlock when its control
button is pressed. (The
docking system can always
be unlocked using the man-
ual release handle).
Electronics fault Reset system by pressing the
control button for 10 seconds.
Contact service
The light on the locking
plate control button flashes
irregularly.
The tension spring is bro-
ken.
Contact service
The wheelchair is not
locked in position in
Permolock C3.
The manual release lever is
in open position.
Move the manual release lever
to the closed position.
The wheelchair does not fit
into Permolock C3 when
driven in over the docking
system.
The height has been incor-
rectly adjusted on
Permolock C3.
Adjust the height of Permolock
C3, see pages 27-29.
The locking mechanism’s
locking arm is loose.
Contact service
The locking mechanism’s
adjustment device is bro-
ken.
Contact service
The spring pins have got
stuck in the chassis.
Contact service