MCO-5AC
Please read the operating instructions carefully before using this product, and keep the operating
instructions for future use.
See page 43 for model number.
Operating Instructions
CO
2
Incubator

Questions and Answers

  • W
    Wendy ChavezAug 3, 2025
    Why is CO gas consumption too high in my Phcbi MCO-5AC?
    • W
      wstaffordAug 3, 2025
      Excessive CO gas consumption in your Phcbi Farm Equipment could be due to: * Frequent opening of the outer and inner doors. * Gas leakage at the connection or pinhole on the gas tube. * An incomplete seal of the inner door gasket. * The access port at the upper left corner being open.
  • L
    Levi ColeAug 7, 2025
    How to troubleshoot a Phcbi MCO-5AC Farm Equipment that does not operate at all?
    • R
      rebeccajacksonAug 8, 2025
      If your Phcbi Farm Equipment isn't operating, check the following: * Ensure the unit is correctly plugged into a power outlet. * Verify that the circuit breaker at the power source is active and that there isn't a power failure. * Confirm that the power supply removable cord is connected to the port on the rear of the cabinet.
  • Y
    yesenialongAug 11, 2025
    What to do if CO gas density doesn't coincide with the set value in Phcbi Farm Equipment?
    • V
      vmorganAug 11, 2025
      If the CO gas density in your Phcbi Farm Equipment doesn't match the set value: * Ensure the secondary pressure is set to 0.03 MPa(G) (0.3 kgf/cm2(G), 4.4 psi(G)). * Check if the gas tube is clogged or chinked. * Verify the humidifying pan is filled with sterile distilled water.
  • M
    Mary GatesAug 15, 2025
    What to do if the chamber temperature is not equal to the set temperature in Phcbi MCO-5AC?
    • T
      Teresa SalazarAug 16, 2025
      If the chamber temperature in your Phcbi Farm Equipment isn't reaching the set temperature: * Ensure the ambient temperature is at least 5 °C less than the set temperature. * Make sure the outer door is closed while the inner door is also closed.
  • D
    Daniel BurnettAug 20, 2025
    Why does it take so long to recover the CO gas density in my Phcbi Farm Equipment?
    • B
      Bailey StephensonAug 20, 2025
      If recovery of CO gas density in your Phcbi Farm Equipment is slow: * A HEPA filter is provided in the CO gas piping. * If CO gas pressure is normal, dust on the HEPA filter might be obstructing the CO gas flow.
  • J
    Jason SwansonAug 25, 2025
    What to do if the chamber humidity does not rise in Phcbi Farm Equipment?
    • C
      Courtney RodriguezAug 25, 2025
      If the chamber humidity in your Phcbi Farm Equipment isn't rising, ensure the humidifying pan is filled with sterile distilled water.
  • C
    Courtney McculloughAug 28, 2025
    Why is the key operation disabled on my Phcbi Farm Equipment?
    • T
      Teresa JonesAug 28, 2025
      If the key operation is disabled on your Phcbi Farm Equipment, the key lock function is likely set to the ON mode.

Summary

Precautions for Safe Operation

General Safety Warnings and Symbols

Explains warning/caution signs and key safety alerts regarding CO2 gas.

Installation and Electrical Safety

Covers safe installation practices, grounding, and electrical connections.

Handling, Storage, and Disposal

Details safe handling, power cord usage, unplugging, and storage guidelines.

Operational and Environmental Safety

Outlines safety for operation, environment, and use of dedicated circuits.

Labels on Incubator

Environmental Conditions

Incubator Components Overview

Control Panel and Keypad

Remote Alarm Terminal

Installation Site Requirements

Installation Procedure

Connecting a CO2 Gas Cylinder

Preventing Contamination

Precautions for Cultures

Initial Cleaning Method

Removing Inner Attachments

Installing Inner Attachments

Filling Humidifying Pan

Water Level Sensor Function

Routine Maintenance

Correct Operation Startup

Setting Chamber Temperature and CO2 Density

Setting Alarm Resume Time

Operation After Power Failure

Describes how the incubator resumes operation after a power interruption.

Key Lock Function

Automatic CO2 Cylinder Changeover

UV Lamp System

Precautions for UV Lamp Use

Essential safety guidelines for operating the UV lamp to prevent harm.

Change UV Lamp ON Period Setting

Procedure to adjust the duration the UV lamp remains active after door closure.

Lighting UV Lamp for 24 Hours

Instructions for activating the UV lamp for continuous 24-hour sterilization.

Alarms, Safety, and Self-Diagnosis

Temperature Calibration

CO2 Calibration

Troubleshooting Guide

Disposing of the CO2 Incubator

Stacking Incubators Safely

Technical Specifications

Performance Characteristics

Service Safety Check Sheet

Phcbi MCO-5AC Specifications

General IconGeneral
BrandPhcbi
ModelMCO-5AC
CategoryFarm Equipment
LanguageEnglish