Separation distance according to
frequency of transmitter (m)
Rated
maximum
output
power of
transmitter
(W)
150 kHz to
80 MHz
d = 1.2 √ P
80 MHz
to 800
MHz
d = 1.2 √ P
800 MHz
to 2.5 GHz
d = 2.3
0.01
0.12 0.12 0.23
0.1 0.37
0.37
0.74
1 1.2 1.2 2.3
10
3.7
3.7 7.4
100
12
12 23
For transmitters rated at a maximum output power
not listed above, the recommended separation
distance d in metres (m) can be estimated using
the equation applicable to the frequency of the
transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation
distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all
situations. Electromagnetic propagation is aected
by absorption and reection from structures,
objects and people.
Explanation of symbols
The warning signs and symbols are essential
to ensure that you use this product safely and
correctly and to protect you and others from injury.
Below you nd the meaning of the warning signs
and symbols on the label and in the user manual.
Read the user manual before you use the
device.
This symbol means that the part of the
device that comes into physical contact
with the user (also known as the applied
part, probe tip) is of type BF (Body
Floating) according to IEC 60601-1.
Indicates the manufacturer.
Symbol for the 'Bluetooth mark'. The
device uses Bluetooth for communication.
Indicates the manufacturer's serial number
so that a specic medical device can be
identied.
Indicates manufacturer's catalog number
of the device.
Field strengths from xed RF transmitters, as
determined by an electromagnetic site survey (a),
should be less than the compliance level in each
frequency range (b).
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all
situations. Electromagnetic propagation is aected
by absorption and reection from structures,
objects and people.
(a) Field strengths from xed transmitters, such
as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due
to xed RF transmitters, an electromagnetic site
survey should be considered. If the measured
eld strength in the location in which the device is
used exceeds the applicable RF compliance level
above, the device should be observed to verify
normal operation. If abnormal performance is
observed, additional measures may be necessary,
such as re-orienting or relocating the device.
(b) Over the frequency range 150 kHz to 80 MHz,
eld strengths should be less than 3V/m.
Table 6 Recommended separation
distances between portable and mobile
RF communications equipment and the
ME equipment or ME system – for ME
equipment and ME systems that are not
life supporting
Recommended separation distances between
portable and mobile RF communications
equipment and the device.
The device is intended for use in an
electromagnetic environment in which radiated
RF disturbances are controlled. The customer
or the user of the device can help prevent
electromagnetic interference by maintaining a
minimum distance between portable and mobile
RF communications equipment (transmitters)
and the device as recommended below,
according to the maximum output power of the
communications equipment.
Elec-
tro sta-
t ic dis-
charge
(ESD)
IEC
61000-
4-2
±6 kV
contact
±8 kV air
±6 kV
contact
±8 kV
air
Floors should be
wood, concrete
or ceramic tile. If
oors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Power
fre-
quen cy
(50/60-
Hz)
mag-
net ic
eld IEC
61000-
4-8
3A/m 3A/m
Power frequency
magnetic elds
should be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.
Table 4 Guidance and manufacturer‘s
declaration – electromagnetic immunity
–for ME equipment and ME systems that
are not life supporting
Guidance and manufacturer’s declaration –
electromagnetic immunity .The device is intended
for use in the electromagnetic environment
specied below. The customer or the user of the
device should assure that it is used in such an
environment.
IMMUNITY
test
- IEC 60601
TEST LEVEL
Compliance
level
Conducted RF
IEC 61000-
4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
Radiated RF
IEC 61000-
4-3
3 V/m
80 MHz to 2.5
GHz
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1.2
√P
d = 1.2 √P 80 MHz to 800MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Electromagnetic emissions and
immunity
The device is approved according to EMC safety
standard IEC 60601-1-2. It is designed to be used
in typical domestic environments.
Table 1 Guidance and manufacturer‘s
declaration – electromagnetic emissions
- for all ME equipment and ME systems
Guidance and manufacturer’s declaration –
electromagnetic emissions
The device is intended for use in the
electromagnetic environment specied below.
The customer or the user of the device should
assure that it is used in such an environment.
Emissions
test
Com pli-
ance
Electromagnetic
environment -
guidance
RF emissions
CISPR 11
Group 1 The DL8740 uses
RF energy only for
its internal function.
Therefore, its RF
emissions are very
low and are not
likely to cause any
interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic
emissions IEC
61000-3-2
Not
applica-
ble
Voltage
uc tu a-
tions/ick er
emissions IEC
61000-3-3
Not
applica-
ble
Guidance and manufacturer‘s declaration
– electromagnetic immunity – for all
ME equipment and ME systems
Guidance and manufacturer’s declaration –
electromagnetic immunity
The device is intended for use in the
electromagnetic environment specied below.
The customer or the user of the device should
assure that it is used in such an environment.
Immu ni-
ty test
IEC
60601
test
level
Com-
pli ance
level
Electromagnetic
environment -
guidance