Technical data
PDM radiology
Table 10.1 PDM radiology technical data
Characteristic Measure
Operational Quantity* Hp(10)
Reproducibility 10% or 1 µSv, whatever is greatest
Dose Rate Range 40 µSv/h – 500 mSv/h
Energy dependence X-/g-rays +/- 30% within N40 – N120
Angular dependence +/- 5% within +/- 5°
+/- 30% within +/- 50°
+200%/-100% within +/- 90°
Temperature dependence +/- 5% within 20-26°C
+/- 25% within 15-35°C
Battery voltage dependence +/- 2% from fully charged until low battery
shutdown
Response time Less than 1s above 100µSv/h, less than 5s
otherwise
Position on body** On torso outside lead apron
*) Hp(10): Personal dose equivalent at a depth of 10 mm according to ISO 4037.
**) Position on body: The Hp(10) measurement is only valid for a position on the torso
outside any lead apron or other protection. (To estimate effective dose to a user, one
must use other means to estimate things like the environmental radiation situation and the
effectiveness of protection.)
10
10.1
DoseAware Base Station Package Document version 1.2
Technical data
10-1
P h i l i p s H e a l t h c a r e 9 8 9 6 0 0 2 1 6 6 2 3
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
All manuals and user guides at all-guides.com

Table of Contents

Related product manuals