Maintenance
Calibrating the Base Station screen
If the touch screen does not respond correctly to user interaction, a touch
screen re-calibration may be needed.
Follow the instructions below to perform a re-calibration:
1 Tap and hold on the Philips start up screen that appears during Base Station
startup until the Setup view appears.
2 Follow the instruction on the bottom of the screen and the subsequent
instructions in the next views.
Cleaning and disinfection
Cleaning and disinfection of this product is required periodically.
Guidelines for each are given below.
Always isolate the equipment from the mains electrical supply before cleaning,
disinfecting or sterilizing to prevent electric shocks.
Never allow water or other liquids to leak into the equipment as this may cause
electrical short-circuits or metal corrosion.
Cleaning and disinfection techniques for both the product and the room
must comply with all applicable local laws and regulations.
Cleaning
Enameled parts and aluminum surfaces should only be wiped clean with a
damp cloth and a mild detergent, and then rubbed down with a dry woolen
cloth. Never use corrosive cleaning agents, solvents, abrasive detergents or
abrasive polishes. If you are not sure about the properties of a cleaning
agent, do not use it.
7
7.1
7.2
W A R N I N G
C A U T I O N
7.2.1
DoseAware Base Station Package Document version 1.2
Maintenance
7-1
P h i l i p s H e a l t h c a r e 9 8 9 6 0 0 2 1 6 6 2 3
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
All manuals and user guides at all-guides.com
all-guides.com

Table of Contents

Related product manuals