- 59 -
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity – This device is
intended for use in the electromagnetic environment specied below. The user of this device
should make sure it is used in such an environment.
immuniTy TEsT iEC 60601 TEsT lEvEl ComplianCE lEvEl ElECTromagnETiC
EnvironmEnT -
guidanCE
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, &
±15 kV air discharges
±8 kV contact
±2 kV, ±4 kV, ±8 kV, &
±15 kV air discharges
Floors should be wood,
concrete or ceramic
tile. If oors are
covered with synthetic
material, the relative
humidity should be at
least 35%.
Electrical fast Transient/
burst
IEC 61000-4-4
±2 kV for power supply
lines, 100 kHz repetition
rate
±2 kV for power supply
lines, 100 kHz repetition
rate
Mains power quality
should be that of
a typical home or
hospital environment.
±1 kV for input-output
lines; 100 kHz repetition
rate
N/A- Device does not
have user I/O lines that
are longer than 3m in
length.
Surge
IEC 61000-4-5
±1 kV dierential mode
±2 kV common mode
±1 kV dierential mode
N/A: Device is Class 2
and does not have earth
connection.
Mains power quality
should be that of
a typical home or
hospital environment.