General Guidelines
14-2 EarlyVue VS30 Instructions for Use
General Guidelines
Sterile accessories are marked with “Sterile” and “Do not use if package is damaged”
symbols. Follow guidelines in the Instructions for Use that come with accessories for how
to handle sterile accessories with damaged packaging.
Always inspect the monitors and accessories for damage and deterioration before each
use.
It is the responsibility of the hospital to determine the appropriate cleaning protocol for
each care unit.
Keep the monitor, cables, and accessories free of dust and dirt. After cleaning and
disinfecting, check the equipment carefully. Do not use the equipment if you see signs of
deterioration or damage.
If you need to return any equipment to Philips, clean and disinfect it first, following the
instructions in this chapter.
Follow these general precautions:
• Always dilute cleaning agents according to the manufacturer’s instructions or use the
lowest possible concentration.
• Do not allow liquid to enter the case.
• Do not immerse any part of the equipment in liquid.
• Do not pour liquid onto the system.
• Never use abrasive material (such as steel wool or silver polish).
• Do not autoclave, steam sterilize, or ultrasonically clean the monitor or cables.
• Do not use bleach on electrical contacts or connectors.
• Do not use alcohol on the patient cables. Alcohol can cause the plastic to become
brittle and fail prematurely.
Caution
If you spill liquid on the exterior of the monitor, use a clean cloth to dry the monitor. If you believe
the liquid may have entered the monitor, contact your biomedical engineer, who can verify the
performance and safety of the equipment.