Physiological Alarms
A-2 EarlyVue VS30 Instructions for Use
NBP(S) Low Medium The Systolic NBP value has dropped below the low
alarm limit.
Pulse Rate Alarms
Pulse High Medium The Pulse Rate value has exceeded the high alarm
limit.
Pulse Low Medium The Pulse Rate value has dropped below the low
alarm limit.
SpO
2
Alarms
SpO2 Desat High The oxygen saturation value has dropped dangerously
low. The Desat alarm sounds when the SpO
2
value is
10 less than the current low limit for adults and
pediatric patients and 5 less for neonates.
SpO2 High Medium The SpO
2
value has exceeded the high alarm limit.
SpO2 Low Medium The SpO
2
value has dropped below the low alarm limit.
Temperature Alarms
Temp High Medium The temperature value has exceeded the high alarm
limit.
Temp Low Medium The temperature value has dropped below the low
alarm limit.
Alarm Message Priority Cause