Auto Set Alarm Limits
A-12 EarlyVue VS30 Instructions for Use
Pulse Rate
(SpO2)
Pulse – 30 (within 80 – 100/min) Pulse + 40 (within 180 – 210/min)
Respiration
(CO2)
awRR – 50 (within 25 – 40) awRR + 25 (within 75 – 95)
SpO2 Same as default alarm limit Same as default alarm limit
Temperature Temp – 0.9°F
Temp – 0.5°C
Temp + 0.9°F
Temp + 0.5°C
Parameter Neonatal Lower Limit Neonatal Upper Limit