EarlyVue VS30 Instructions for Use B-3
Guidance and Manufacturer’s Declaration
Guidance and Manufacturer’s Declaration
The EarlyVue VS30 vital signs monitor is intended for use in the electromagnetic
environment specified in the following tables. The customer or the user of the product
should ensure that it is used in such an environment.
Table B-1 Electromagnetic Emissions
Table B-2 Enclosure Port
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1, Class B The EarlyVue VS30 vital signs monitor is
suitable for use in a professional
healthcare facility environment.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Immunity Test
Immunity Test
Level
Compliance
Level
Electromagnetic Environment
- Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2, ±4,±8, ±15 kV air
±8 kV contact
±2, ±4,±8, ±15 kV air
The EarlyVue VS30 vital signs
monitor is suitable for use in a
professional healthcare facility
environment.
Radiated RF
Electromagnetic
Field
IEC 61000-4-3
3 V/m
80 MHz - 2.7 GHz
3 V/m
80 MHz - 2.7 GHz
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
Refer to table B-3 Refer to table B-3
Power frequency
magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz