EarlyVue VS30 Instructions for Use B-5
Guidance and Manufacturer’s Declaration
Table B-4 Input a.c. Power Port
Immunity Test Immunity Test Level Compliance Level
Electromagnetic
Environment - Guid-
ance
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV ± 2 kV The EarlyVue VS30
vital signs monitor is
suitable for use in a
professional
healthcare facility
environment.
Surge
Line to line
IEC 61000-4-5
± 0.5 kV, ±1 kV ± 0.5 kV, ±1 kV
Surge
Line to ground
IEC 61000-4-5
± 0.5 kV, ±1 kV, ±2 kV ± 0.5 kV, ±1 kV, ±2 kV
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3 V
0.15 MHz – 80 MHz
6 V in ISM bands
between 0.15 MHz and
80 MHz
3 V
0.15 MHz – 80 MHz
6 V in ISM bands
between
0.15 MHz and 80 MHz
Voltage dips
IEC 61000-4-11
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°;
0% UT; 1 cycle and
70% UT; 25/30 cycles
Single phase: at 0°
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°;
0% UT; 1 cycle and
70% UT; 25/30 cycles
Single phase: at 0°
Voltage interruptions
IEC 61000-4-11
0% UT; 250/300 cycle 0% UT; 250/300 cycle