EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
Module 4 Preventive Maintenance
Preventive maintenance procedures are very important in a clinical
setting. Be sure to follow the health care facilities documentation
procedures. Philips recommends that preventive maintenance
procedures be performed every 12 months. The following equipment
is needed for basic preventive maintence:
System cart lift tool
Molygraph SL3144 bearing grease
Soft microfiber or lint-free cloth
Liquid screen cleaner for LCDs
FSE Tool Kit
Vacuum
System Backup
System backup is very important. Backup the system anytime you are
in front of it! Your backup is not complete without printouts!
To create backup on EPIQ systems:
Support System Management Backup/Restore
Support System Management System Information
Setup About Additional License Options
Setup About Licensed Transducers
You can also backup to removable media such as DVD. Supported
removable media includes DVD, CD and USB storage devices. Some
settings can be imported via removeable media to other systems of
the same model:
Analysis
Annotation/ body markers
Audio/ Video
DICOM
DNS
Note: The systems TCP/IP
address is not backed up!
This is why the printouts
are so important.