4 Cableless Monitoring
105
After Monitoring
1 After monitoring is done, remove the CL F&M Pod from the electrode patch.
2 Clean and disinfect the CL F&M Pod, see “Care and Cleaning” on page 261.
3 Dock the CL F&M Pod back at the base station.
4 Gently remove the electrode patch and the electrodes from the abdomen of the patient.
5 Dispose of the electrode patch.
6 Remove the residuals of the adhesive from the patient's skin.
CL Wide Range Pod
The CL Wide Range Pod extends the signal range for the cableless measurement devices worn by the
ambulating patient.
1 Assign the CL Wide Range Pod at the base station as described in “CL Pod Assignment” on
page 99.
2 Activate the CL Wide Range Pod by taking it off the base station.
The CL Wide Range Pod connects itself over WLAN and LAN with the fetal monitor. As soon as
the connection to the monitor is established, the CL Wide Range Pod takes over from the base
station, and sends the signals from the cableless devices to the monitor. The connection status of
the CL Wide Range Pod is shown on the monitor, see “Cableless Status Indication” on page 95.
3 Hand the CL Wide Range Pod to the patient to carry with her, or optionally wearing it using the
Mobile CL Transmitter Cradle (989803168881). The CL Wide Range Pod should be worn with its
rear side facing the body.
CL Wide Range Pod carried in hand, or CL Wide Range Pod worn with the belt clip (supporting
loop and belt clip part of the optional Mobile CL Transmitter Cradle).
4 While the CL Wide Range Pod is active, cableless devices can still be assigned or unassigned at the
base station dynamically to the patient. To hand the signal transfer back from the CL Wide Range
Pod to the base station, redock the CL Wide Range Pod at the base station.