7 Patient Alarms and INOPs
130
*** Brady (Pulse)
***Brady/P xxx<yyy
SpO
2
The heart rate from the Pulse signal has
fallen below the bradycardia limit.
xxx denotes the lowest measured value; yyy
is the bradycardia limit.
Numeric flashes, red alarm
message, alarm tone.
*** Desat
*** Desat xx < yy
SpO
2
The SpO
2
value has fallen below the
desaturation alarm limit. xx denotes the
lowest measured value, and yy is the
desaturation limit.
Numeric flashes, red alarm
message, alarm tone.
** dFHR1 High
** dFHR2 High
** dFHR3 High
**dFHR1 xxx>yyy
**dFHR2 xxx>yyy
**dFHR3 xxx>yyy
FHR (DECG) The fetal heart rate obtained from DECG
has risen above the high alarm limit.
xxx denotes the highest measured value, and
yyy is the high alarm limit.
Numeric flashes and high
limit is highlighted, yellow
alarm message, alarm tone.
** dFHR1 Low
** dFHR2 Low
** dFHR3 Low
**dFHR1 xxx<yyy
**dFHR2 xxx<yyy
**dFHR3 xxx<yyy
FHR (DECG) The fetal heart rate obtained from DECG
has fallen below the low alarm limit.
xxx denotes the lowest measured value, and
yyy is the low alarm limit.
Numeric flashes and low limit
is highlighted, yellow alarm
message, alarm tone.
*** Extreme Brady
***xBrady xxx<yyy
MECG The maternal heart rate obtained from the
maternal ECG has fallen below the extreme
bradycardia limit.
xxx denotes the lowest measured value, and
yyy is the extreme bradycardia limit.
Numeric flashes, red alarm
message, alarm tone.
*** Extreme Tachy
***xTachy xxx>yyy
MECG The maternal heart rate obtained from the
maternal ECG has risen above the extreme
tachycardia limit.
xxx denotes the highest measured value, and
yy is the extreme tachycardia limit.
Numeric flashes, red alarm
message, alarm tone.
** FHR1 High
** FHR2 High
** FHR3 High
**FHR1 xxx>yyy
**FHR2 xxx>yyy
**FHR3 xxx>yyy
FHR
(ultrasound)
The fetal heart rate obtained from
ultrasound has risen above the high alarm
limit.
xxx denotes the highest measured value, and
yyy is the high alarm limit.
Numeric flashes and high
limit is highlighted, yellow
alarm message, alarm tone.
** FHR1 Low
** FHR2 Low
** FHR3 Low
**FHR1 xxx<yyy
**FHR2 xxx<yyy
**FHR3 xxx<yyy
FHR
(ultrasound)
The fetal heart rate obtained from
ultrasound has fallen below the low alarm
limit.
xxx denotes the lowest measured value, and
yyy is the low alarm limit.
Numeric flashes and low limit
is highlighted, yellow alarm
message, alarm tone.