4 Cableless Monitoring
99
NOTE
When a CL transducer is unassigned, all the measurements from this transducer are no longer
monitored; e.g. if you unassign a CL Toco
+
MP transducer, the Toco measurement, maternal Pulse, or
possibly also MECG or DECG are no longer monitored.
CL Pod Assignment
The CL Pods are assigned by simply docking them at the Avalon CL base station connected to the
patient’s fetal monitor.
CL Pod Unassignment
The CL Pods have to be unassigned:
• before assigning them to another base station or to another patient
• to allow cleaning, before docking them back onto the base station
There are three methods to do this:
1 Open the setup menu of the measurement e.g. by selecting the numeric on the screen.
2 Select Remove from the open setup menu.
3 A confirmation window opens with the prompt Confirm and Device will be removed.
4 Select the Confirm pop-up key. A message confirms <cl Measurement> has been removed.
or
1 Open the Tele Info window.
2 Select the symbol of the CL Pod, and select Remove from the pop-up menu.
3 A confirmation window opens with the prompt Confirm and Device will be removed.
4 Select the Confirm pop-up key. A message confirms <cl Measurement> has been removed.
or
1 Redock the CL Pod at the base station.
2 Switch the base station to stand-by.
3 Remove the CL Pod; it is now unassigned.
To unassign a IntelliVue CL Pod using its user interface, see the Cableless Measurement
Instructions for Use.
NOTE
If you unassign a CL Pod, all the measurements from it are no longer monitored, or are monitored by
a connected lesser prioritized sensor.