Technical data EN13544-1
MMAD 3.99 μm
Aerosol output 1.19 mL
Aerosol output rate 0.26 mL/min
Max fill volume 8 ml
Max medication temperature increase at max fill <10
o
C above ambient temperature
Noise level <35dB at 1m
Distribution of particles >5 μm 35.6%
Distribution of particles in range 2 - 5 μm 52.3%
Distribution of particles in range <2 μm 12.1%
Respirable fraction (%<5 μm) 64.4%
Percentage of fill volume (2.5 mL) as aerosol 10.2%
output (delivered to filter) in one minute
Residual volume 0.30 mL
If the device has been stored at the extremes of the storage temperature, please allow at least 1 hour at room
temperature before using the device.
Use of a face mask may affect the device performance. Data available on request.