9
Mask cleaning
• If using a mask and mask adapter, once a week, agitate in warm soapy water for 2 minutes or
soak for 10 minutes, rinse and air dry.
Maintenance
Mouthpiece assembly can be cleaned (boiled/disinfected) up to 52 times over 12 months.
To keep your InnoSpire Go working at the optimum level you should replace the mouthpiece
assembly and mask, if used, every 12 months as this is a consumable part.
This is based on a typical usage profile of 3 treatments per day.
LED Indication
LED Indication What it means
SOLID GREEN The battery is charged
SOLID AMBER There is enough charge for at least one
more treatment
FLASHES AMBER FIVE TIMES There is not enough charge to take a treatment
FLASHES GREEN OR AMBER FIVE TIMES The treatment is finished
PULSES GREEN The battery is being charged
1128271R02_InnoSpireGo_Euro2_Spreads_IFU_10B_FINAL.indd 9 11/8/17 5:07 PM