44
4.2.3.
Vde,mV
and
Kl
>
preset
part
test
SET
INSTRUMENT
IN
POSITION
V==
OR
Vex.
a(]>
PRESET
CONNECT
1V
TO
THE
V@mA_
AND
0
SOCKET
CORRECT
(90mV)
VOLTAGE
ON
THE
JUNCTION
OF
R1209
AND
R1210
?
CHECK
V1100
AND
V1101
CONNECT
10V
TO
THE
vamA
AND
THE
0
SOCKET
CORRECT VOLTAGE
(900mV)
ON
THE
JUNCTION
OF
R1209
AND
1210
?
CHECK
V1100
AND
V1101
RNGE
AND
RNGO
SET
THE
INSTRUMENT
IN
POSITION
mV==
CONNECT
100mV
TO
THE
VamA
AND
0
SOCKET
CORRECT VOLTAGE
(90mv}
ON
THE
JUNCTION
OF
R1209
AND
R1210?
CHECK
SWITCH
PCB
ST4422
840629
NOTE:
Measurement
zero
is
the
low
socket.
11,11
11,11
1111,11
4411,11
NOTE:
The
voltage
on
the
junctions
must
be
measured
with
a
high
impedance
voltmeter
(100
MQ).
If
this
is
not
possible
take
account
of
the
shunting
effect
finput
impe-
dance
PM
2519).