19User Manual
Specications
Environmental
Operating Temperature: 5° to 35° C (41° to 95° F)
Storage Temperature: -20° to 60° C (-4° to 140° F)
Relative Humidity (operating & storage): 15 to 95% (non-condensing)
Atmospheric Pressure: 101 to 77 kPa (0 - 2286 m / 0 - 7500 ft)
Physical
Dimensions: 18 x 14 x 10 cm (7” L x 5.5” W x 4” H)
Weight (Device with power supply): Approximately 1.53 kg (3.37 lbs)
Standards Compliance This device is designed to conform to the following standards:
IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment
EN ISO 17510-1 Sleep Apnea Breathing Therapy Devices
EN 60601-1-2 Electromagnetic Compatibility
RTCA/DO-160F section 21, category M; Emission of Radio Frequency Energy
IEC 60601-1 Classication
Type of Protection Against Electric Shock: Class II Equipment
Degree of Protection Against Electric Shock: Type BF Applied Part
Degree of Protection against Ingress of Water:
Device: Drip Proof, IP22
60W power supply: Drip Proof, IP22
80W power supply: Drip Proof, IP22
Mode of Operation: Continuous
Electrical
AC Power Consumption (with 60W power supply): 100 – 240 VAC, 50/60 Hz, 2.1 A
AC Power Consumption (with 80W power supply): 100 – 240 VAC, 50/60 Hz, 2.0 A
DC Power Consumption: 12 VDC, 6.67 A
Fuses: There are no user-replaceable fuses.
Declared Dual-Number Noise Emissions Values In accordance with ISO 4871
The measured A-weighted emission sound pressure level is 27 dB(A) with an uncertainty of 2 dB(A).
The measured A-weighted sound power level is 35 dB(A) with an uncertainty of 2 dB(A).
Notes:
• These measurements apply to this device with an optional humidier. Use of this device without a
humidier would result in measurements equal to or less than the stated values.
• Values determined according to noise test code given in ISO 17510-1:2007, using the basic standards
ISO 3744 and ISO 4871.