53
Immunity
Test
IEC60601-1-2
Test Level
Compliance Level
Electrostatic
Discharge
(ESD) IEC
61000-4-2
+/- 8 kV contact
+/- 2 kV , +/- 4 kV,
+/- 8 kV, +/- 15 kV air
+/- 8 kV contact
+/- 2 kV , +/- 4 kV,
+/- 8 kV, +/- 15 kV air
Radiated RF
Electromag-
netic Field
Immunity
Test IEC
61000-4-3
10 V/m 80 MHz – 2.7 GHz
80% AM at 1 kHz
10 V/m 80 MHz – 2.7 GHz
80% AM at 1 kHz
Proximity
Magnetic
Field IEC
61000-4-3
9 V/m; 710, 745, 780 MHz;
Pulse modulation 18 Hz
9 V/m;
5240, 5500, 5785 MHz;
Pulse modulation 217 Hz
27 V/m; 385 MHz; Pulse
modulation 18 Hz
28 V/m; 450 MHz FM +/- 5
kHz deviation; 1 kHz sine
28 V/m; 810, 870, 930 MHz;
Pulse modulation 18 Hz
28 V/m; 1720, 1845,
1970, 2450 MHz;
Pulse modulation 217 Hz
9 V/m; 710, 745, 780 MHz;
Pulse modulation 18 Hz
9 V/m;
5240, 5500, 5785 MHz;
Pulse modulation 217 Hz
27 V/m; 385 MHz; Pulse
modulation 18 Hz
28 V/m; 450 MHz FM +/- 5
kHz deviation; 1 kHz sine
28 V/m; 810, 870, 930 MHz;
Pulse modulation 18 Hz
28 V/m; 1720, 1845,
1970, 2450 MHz;
Pulse modulation 217 Hz
Electromagnetic immunity
Immunity
Test
IEC60601-1-2
Test Level
Compliance Level
Rated power
frequency
magnetic
elds IEC
61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Enclosure
Port
Immunity to
Proximity
Magnetic
Fields IEC
61000-4-39
Test frequency 30kHz,
modulation: CW,
Immunity test level 8
Test frequency
134, 2 kHz,
Pulse modulation: 2.1 kHz,
Immunity test level 65
Test frequency 13.56 kHz,
Pulse modulation: 50kHz,
Immunity test level 7.5
Test frequency 30kHz,
modulation: CW,
Immunity test level 8
Test frequency
134, 2 kHz,
Pulse modulation: 2.1 kHz,
Immunity test level 65
Test frequency 13.56 kHz,
Pulse modulation: 50kHz,
Immunity test level 7.5