English
51
No. Remark Indications Description Frequency/
Hz
Pulse
width/µs
Timer/
min
10
Gate
Control
dynamic 1
Relaxing of
muscles
Intensity in 1 seconds
on maximum, within 1
seconds to zero. Both
channels in alternating
run.
80 150 30
11
Gate
Control
dynamic 2
Relaxing of
muscles
Intensity in 0,25
seconds on maximum,
within 0,25 seconds to
zero. Both channels in
alternating run.
80 150 30
12
Deep- TENS Muscular-related
pain, deep pain
causes
Pulse blocks of 4 pulses,
the single pulses have a
distance of 200 µs
100 75 30
Modifiable Programmes from Series C
13
Modifiable
TENS
stimulation
Pain therapy as in
Programme 1-3
Both channels equal 1-120 75-300 10-90
14
Modifiable
muscle
stimulation
Muscular-related
pain, deep pain
causes
Rising time 2 seconds,
working time 5 seconds,
falling time 1 seconds,
pause time 12 seconds,
automatic intensity
adjustment
1-120 75-300 10-90
15
Modifiable
muscle
relaxing
Relaxing of
muscles
Intensity in 1 seconds
on maximum, within 5
seconds to zero. Both
channels in alternating
run
1-120 75-300 10-90
Modifiable Programme from Series E
16
Mono-
phase
programme
Nerve block
acc. to Jenkner
(may cause skin
reddening)
Both channels equal 1-120 75-300 20