Operation, interfaces and standards
S12C
S16
Touch screen
operation
5.7 inch (capacitive)
Aspect ratio 4:3
(640 x 480 pixels)
4.3 inch (resistive)
Aspect ratio 5:3
(800 x 480 pixels)
Interfaces
12V, USB, IR, Mini-USB, SD
12V, 4 x USB, IR, DVI
WLAN
WLAN, LAN (RJ-45)
Standards
EN 60601-1
IR
Devices use infrared light with a wavelength of 870nm and a maximum
intensity of 135mW/sr.
WLAN interface standards:
Physical layer
Supports 802.11 b and g standards
Network architecture types
Communication with wired networks via access points
Security
WPA and WPA2
Transmit power
802.11b/g: 50 mW (+17 dBm) typically
Frequency range
2,400 - 2.4897 GHz
Operating channels
1-11 for North America and others,
1-13 for Europe and others
Channels
Supports all channels of 802.11d-APs.
RX sensitivity
-86 dBm typically @ 11 Mbps,
-82 dBm @ 6 Mbps,
-69 dBm @ 54 Mbps
Data rates
802.11g: 6, 9, 12, 18, 24, 36, 48, 54 Mbps
802.11b: 1, 2, 5.5, 11 Mbps
Environmental conditions for operation and storage
According to standard IEC 60601-1-2, we are required to inform you that the device has
been designed for the environment category "Environment in areas of domestic health care,
except for vehicles and planes”. It must be stored in its original packaging. Do not place the
original packaging near heat sources (radiators, fan heaters, etc.). When you remove the
device from its original packaging, do not expose it to sunlight.
Storage
Temperature
0 to +50 °C (+32 to +122 °F)
Humidity
10 to 80 % non-condensing
Operation
Temperature
+10 to +40 °C (+50 to +104 °F)
Humidity
20 to 80 % non-condensing
Max. height
Max. storage height
<2000 m (6,560 ft)
Max.operating
height
<2000 m (6,560 ft)