6
TROUBLESHOOTING
Problem
Nostimula-
tionfelt
Intermittent
stimulation
Not
powerful
enough
Surgesof
power
Possible
Cause
Dead battery
Battery
in
b a c k w a r d s
Lead-off
Broken lead wire
Very
dry
electrode
Defective unit
Loose
connection
unit
or
electrode
Broken lead wire
Electrode
too
dry
or
worn
out
Battery running
low
Electrodes
resistance
too
high
Improper mode
choice
Electrode
peeling
lose
from
skin
Corrective
Action
Replacewithfresh
battery
Remove
and
insert
correctly
Pullouttheunused
lead
plug
from
unit
Replace cord
set
w i t h
anothertocheckfor
broken wire
Re-apply electrodes
contact dealer
Reconnect,
making
sure contact
is
secure
Replace cord
set
w i t h
anothertocheckfor
broken wire
Re-apply electrodes
Replace battery
Uselowimpedance
electrodes
Proper operation
function mode, push
function button
for
another mode
Replace electrode
if
necessary
6
TROUBLESHOOTING
I-
Problem
Stimulation
weakness
minutes
after
start
Unwanted
muscle
twitches
occur
Stimulation
strong,
but
ineffective
Possible
Cause
Normal
adaptation
process
•Amplitude
too
high
Improper
electrode
placement
Corrective
Action
Increase
amplitude
Push
function
button
for
another mode
Decrease amplitude
Push function button
for
another mode
Reposition
electrodes;
consult
your
clinician
Push
function
button
for
another mode
If,
after using this guide,
the
unit does
not
operate properly,
theun
i t
sho uld
be
ser vic ed.
Please call
yourdealer
for
instruction.