How to Change Parameter Settings:
Steps: Hold M+ key power = (Enter parameter setting password)
MF= mode function: 1= counting
3= liter / % ( Optional)
DSS= display mode: 0 = normal mode
1= power save mode.
A-0 = Set time to blank the display in idle condition (0-9 Seconds)
DU-0 = dummy zero: 0= without dummy zero
1= with dummy zero
2= dual accuracy (0.5/1) or (5/10)
CS = Capacity Selection: 1 = Only One Capacity
2= Two Capacity
3= Three Capacity
CAP-1= first capacity ( first scale range for triple capacity) (Enter)
(If not use inter 00000 value)
DP1 = decimal point (DP) of cap1
FD1= accuracy class (FIRST DIGIT) of cap1
CAP-2= second scale capacity (enter)
(If not use inter 00000 value)
DP2 = decimal point (DP) of cap2
FD2 = accuracy class (FIRST DIGIT) of cap2
CAP-3= scale capacity (enter)
DP3 = decimal point (DP) of cap3
FD3 = accuracy class (FIRST DIGIT) of cap3
Litd = Feed liter density
How to see LoadCell counts:
Step 1: Hold MODE key power on = see machine internal count
Step 2: On No-load machine count must be 3000 to 12000 count
Step 3: Usable machine counts are up to 300000