POCTech CT100C (CT10) CGM
Frequently Asked Questions
March 2020 Global FAQ Master
Page 13 of 21
Question
Response
At this time, POCTech CT-10 is a continuous glucose monitoring system indicated for the
management of diabetes in persons age 18 years and older.
50.
Does POCTech CT-10 contain latex?
The POCTech CT-10 Sensor is latex free.
(Reference: CT202 Safety Data Sheet)
51.
What should I do if I experience a skin reaction?
Be sure you are rotating the insertion site. If the problem persists, speak with your healthcare
provider. People with an allergy to adhesives should not use POCTech CT-10.
Agent Note: Complete Potential Adverse Event Report
52.
Can heat affect the how well POCTech CT-10
works?
POCTECH CT-10 should not be exposed to any external heat sources (like a sauna or steam
room) for long periods of time.
During shipment, POCTech CT-10 sensors can be maintained between 2
C 45
C (35
F
113
F) for up to 30 days.
Once you receive the shipment, POCTech CT-10 sensors should be stored away from heat and
direct light at temperatures between 2
C 30
C (35
F 86
F). Keep POCTech CT-10 Sensors
in the individual sealed package until ready for use. When ready for use, allow sensor 30 minutes
to warm to room temperature. During normal wear time, sensor should be maintained between
temperature 15
C 40
C (59
F 104
F).
(Reference: User Guide)
If you think the sensors may have been exposed to temperatures below 2
C(35
F) or above 45
C
(113
F), it is recommended that you do not use that sensor.
53.
Will the Sensors be damaged if left out in hot/cold
weather when shipped by mail?
During shipment, POCTech CT-10 sensors can be maintained between 2
C 45
C (35
F
113
F) for up to 30 days.
Once you receive the shipment, POCTech CT-10 sensors should be stored away from heat and
direct light at temperatures between 2
C 30
C (35
F 86
F). Keep POCTech CT-10 Sensors
in the individual sealed package until ready for use. When ready for use, allow sensor 30 minutes
to warm to room temperature. During normal wear time, sensor should be maintained between
temperature 15
C 40
C (59
F 104
F).
(Reference: User Guide)
If you think the sensors may have been exposed to temperatures below 2
C(35
F) or above 45
C
(113
F), it is recommended that you do not use that sensor.
54.
Can POCTech CT-10 Sensors be inserted into skin
that is sunburned?
Do not apply the Sensor to oily, broken, burned, sunburned, cut, or irritated skin.