M X R F L O W M E T E R
U S E R S M A N U A L
Instructions and Safety Considerations
MXR Models of Porter Conscious Sedation Flowmeters
And Bag Tee (Accessory)
Cabinet Mount Models:
2 05 0 W , 2 0 55 W , 2 0 60 W , 2 0 66W , 20 6 5 W , 2 070 W , 3 0 65 W , 3 0 6 5 W 1 5 , an d 3 066 W
Wall Mount Models:
2000
2000-OS
3000
3000-OS
Bag Tee Models:
P1407A
P2407A
P1407B
CABINET MOUNT MODELS

Questions and Answers

  • E
    Erin RodriguezAug 13, 2025
    What to do if there is no O or N2O flow in my Porter 3000 Medical Equipment?
    • N
      Nicole LewisAug 13, 2025
      If there's no flow of O or N2O when the ON/OFF switch is ON, and the knobs are set, check the following: * Ensure the O regulator in the tank room is turned ON. * Verify the machine is connected to the pipeline system via a wall outlet. * If using O cylinders, replace empty ones with full cylinders.
  • R
    Robin AcostaAug 17, 2025
    Why is the patient not feeling the effects of gases with the Porter 3000?
    • R
      Robin HoffmanAug 17, 2025
      If the patient isn't feeling the effects of the gases, it might be due to the following reasons: * The outer mask may not be properly fit to the patient’s face. Adjust it so the inner mask is secure but the outer mask is slightly off the face. * The gas flows may not be meeting the patient’s requirements. Try readjusting the gas flows to achieve acceptable bag action.
  • H
    hkellerAug 21, 2025
    What to do if the bag is going flat during the procedure with Porter 3000 Medical Equipment?
    • D
      Douglas ChoiAug 21, 2025
      If the bag is going flat during the procedure, consider these possible causes: * The gas flows might not be meeting the patient’s requirements. Readjust the gas flows to achieve acceptable bag action. * The outer mask may not be properly fit to the patient’s face. Ensure the inner mask is secure to the face, while the outer mask remains slightly off the face.
  • M
    Melissa BrownAug 25, 2025
    How to fix no N2O flow but O flow in Porter 3000?
    • M
      melissa14Aug 25, 2025
      If you can get O flow but not N2O flow, check these: * Ensure the N2O cylinder is turned ON. * Replace the N2O cylinder if it's empty with a full one.
  • G
    gutierrezandrewAug 28, 2025
    Why can't I get 9 1/2 L/min O flow with my Porter 3000 Medical Equipment?
    • C
      Cheryl FergusonAug 28, 2025
      If you cannot get 9 ½ L/min O flow with the concentration control knob OFF and the flow control knob full ON, the oxygen pressure setting might be low. Check the O regulator and ensure the O manifold pressure is at 50 PSIG ±2 PSIG. If it's not, call your Dealer for service.
  • K
    kphillipsSep 1, 2025
    Why do both gas flows vary proportionally in my Porter Medical Equipment?
    • M
      Morgan LewisSep 1, 2025
      If, with the N2O concentration set, both flows vary proportionally without a change in the flow setting, the O regulator might be varying the pipeline pressure. Check the O regulator to ensure the O manifold pressure is 50 PSIG ±2 PSIG. If it's not, call your Dealer for service.
  • L
    lcomptonSep 5, 2025
    What to do if gas is leaking from the ON/OFF switch of my Porter Medical Equipment?
    • M
      Michael GilbertSep 5, 2025
      If gas is leaking from the ON/OFF switch, it may be due to a nick or cut in the o’rings on the ON/OFF switch. Call your Dealer for service.
  • K
    Kenneth EstradaSep 9, 2025
    What to do if meter flows N2O without O flow in Porter 3000 Medical Equipment?
    • W
      waynepetersonSep 9, 2025
      If the meter flows N2O without any O flow in the O tube, this indicates a failsafe failure. Take the equipment out of service and return it to Porter.

Summary

Warnings and Precautions

General Safety Warnings and Cautions

Details critical safety warnings, cautions, and operational guidelines for the MXR Flowmeter to ensure safe use.

Crossed Lines Safety and Prevention

Identifying and Avoiding Gas Pipeline Crossings

Explains hazards of crossed gas lines and provides methods to identify and prevent them.

Safety Guidelines and Compliance

Flowmeter Mounting Options

Safety Feature Checks

Nitrous Oxide Failsafe System Check

Procedure for checking the Nitrous Oxide Failsafe System before each use.

Concentration Control Knob Valve Check

Procedure for monthly checking the Concentration Control Knob valve function.

Monthly Operational Checks

Oxygen Flush and Leak Testing

Procedures for monthly checking the Power Oxygen Flush and testing the 3 liter bag/rubber goods for leaks.

Maintenance and Calibration

Monthly Calibration Check Procedure

Details the procedure for performing a monthly calibration check of the flowmeter.

Directions for Use

Basic Delivery Technique

Maintenance and Service

Troubleshooting Chart

Porter 3000 Specifications

General IconGeneral
BrandPorter
Model3000
CategoryMedical Equipment
LanguageEnglish