39
GUIDANCE AND MANUFACTURER’S DECLARATION–
ELECTROMAGNETIC EMISSIONS
PowerDot PD-01 stimulator is intended for use in the electromagnetic
environment specified below. The customer or the user of the PowerDot
PD-01 stimulator should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment–
Guidance
Radiated
Emissions
CISPR 11
Group 1 PowerDot PD-01 stimulator
uses RFenergy only for its
internal function. Therefore, its
RFemissions are very low and are
not likely to cause any interference
in nearby electronic equipment.
Conducted
Emissions
CISPR 11
Not Applicable PowerDot PD-01 does not connect
to AC source directly.