EasyManua.ls Logo

PowerDot PD-01V - Page 40

PowerDot PD-01V
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
GUIDANCE AND MANUFACTURER’S DECLARATION–
ELECTROMAGNETIC IMMUNITY
PowerDot PD-01 device is intended for use in the electromagnetic
environment specified below. The customer or the user of the PowerDot
PD-01 device should assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment –
Guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
±6kV contact
±8kV air
±6kV
contact
±8kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for input/
output lines
Not
Applicable
PowerDot PD-01 does
not connect to AC
source directly.
Surge IEC
61000-4-5
±1kV Line to Line
(or Neutral)
±2kV Line(s) to
earth
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input units
IEC 61000-4-11a
<5% UT (>95%
dip in UT)
for 0.5
Cycle
40% UT (60%
dip in UT) for 5
Cycles
70% UT (30%
dip in UT) for 25