4 9202MCM101
9202 Configuration Parameter List
Category
Parameter
Name
No.
Register
Address
Register
Size
Read/
Write
Type Description Values
GENERAL DEVICE NUMBER 0 0 1 RO
UNSIGNED
INTEGER
Defines the actual
device number
9202 = 37378 (0x9202))
GENERAL DEVICE VERSION 1 1 1 RO
UNSIGNED
INTEGER
Product version 0
GENERAL PASSWORD 3 3 1 R/W
UNSIGNED
INTEGER
Password
for entering
configuration menu
Range: 0...9999
GENERAL PASSWORD TRY 4 4 1 R/W
UNSIGNED
INTEGER
Password attempted
for entering
configuration menu
Range: 0...9999
INPUT
INDICATE SENSOR
SHORTAGE
5 5 1 R/W INTEGER
Indicate sensor
shortage via relay
and/or rail.
NO = 0
YES = 1
INPUT
INDICATE SENSOR
BREAKAGE
6 6 1 R/W INTEGER
Indicate sensor
breakage via relay
and/or rail.
NO = 0
YES = 1
INPUT RAIL ERROR 7 7 1 R/W INTEGER Indicate error on rail.
NO = 0
YES = 1
DISPLAY
DISPLAY
CONTRAST
8 8 1 R/W
UNSIGNED
INTEGER
Contrast in the LCD
display
Range: 0...9
DISPLAY
DISPLAY
BACKLIGHT
9 9 1 R/W
UNSIGNED
INTEGER
Backlight intensity
in LCD
Range: 0...9
DISPLAY LINE FUNCTION 10 10 1 R/W INTEGER
Information shown
in input line of
display in monitor
mode
(normal mode).
TAG = 1
D.OUT = 2
ALTERNATING = 3
GENERAL
ENABLE
PASSWORD
11 11 1 R/W INTEGER
Password protect
entry to configura-
tion menu)
NO = 0
YES = 1
GENERAL
HELP TEXT
LANGUAGE
12 12 1 R/W INTEGER
Language for the
help texts shown on
display.
UK = 0
DK = 1
DE = 2
FR = 3
SE = 4
IT = 5
ES = 6
GENERAL
SIL ENABLE
*1)
13 13 1 RO INTEGER Read SIL mode.
NO = 0
YES = 1
GENERAL DEVICE TYPE 14 17 1 RO INTEGER
Defines the
actual device type
together with the
device no.
(e.g. 9202B1A =
0x0B1A)
B1A = 2842 (0x0B1A)
B1B = 2843 (0x0B1B)
B2A = 2858 (0x0B2A)
B2B = 2859 (0x0B2B)
B3A = 2874 (0x0B3A)
B3B = 2875 (0x0B3B)
INPUT
CHANNEL 1
FUNCTION
15 18 1 R/W
UNSIGNED
INTEGER
Select function type
for channel 1.
INVERT = 0
DIRECT = 1
DISPLAY
CHANNEL 1
TAG TEXT
16 20 3 R/W ASCII CHAR
Tag of the device
channel 1
(5 characters).
Range: ASCII values from 32 to 90 (‘ ‘ to ‘Z’)
INPUT
CHANNEL 2
FUNCTION
17 23 1 R/W
UNSIGNED
INTEGER
Select function type
for channel 2.
INVERT = 0
DIRECT = 1
DISPLAY
CHANNEL 2
TAG TEXT
18 25 3 R/W ASCII CHAR
Tag of the device
channel 2
(5 characters).
Range: ASCII values from 32 to 90 (‘ ‘ to ‘Z’)
GENERAL
Configuration
counter
101 101 1 RO
UNSIGNED
INTEGER
This counter will
count the number
of times the
configuration has
been changed.
The counter is reset
on power-up
Range 0...65536
*1) If SIL mode is set by an authorized SIL user (SIL ENABLE = YES), all parameters become READ ONLY