EasyManuals Logo

PR 5714 User Manual

PR 5714
25 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
5714V103-UK 11
Voltage input:
Measurement range ................................................ 0...12 VDC
Programmable measurement ranges ............... 0...1 / 0.2...1 /
0...10 / 2...10 VDC
Input resistance........................................................ Nom. 10 MΩ
Outputs:
Display:
Display readout......................................................... -1999...9999 (4 digits)
Decimal point ............................................................. Programmable
Digit height ................................................................ 13.8 mm
Display updating ...................................................... 2.2 times / s
Input outside input range is
indicated by................................................................ Explanatory text
Current output:
Signal range (span) ................................................. 0...20 mA
Programmable signal ranges ............................... 0...20 / 4...20 /
20...0 / 20...4 mA
Load (max.) ................................................................. 20 mA / 800 Ω / 16 VDC
Load stability ............................................................. 0.01% of span / 100 Ω
Sensor error detection ........................................... 0 / 3.5 / 23 mA / none &
NAMUR NE 43 up / downscale ........................... 23 mA / 3.5 mA
Output limitation:
on 4...20 and 20...4 mA signals .................. 3,8...20.5 mA
on 0...20 and 20...0 mA signals .................. 0...20.5 mA
Current limit ............................................................... 28 mA
Relay outputs:
Relay function ........................................................... Setpoint
Hysteresis ................................................................... 0...100%
On and Off delay ...................................................... 0...3600 s
Sensor error detection ........................................... Make / Break / Hold
Max. voltage .............................................................. 250 VRMS
Max. current ............................................................... 2 A / AC
Max. AC power .......................................................... 500 VA
Max. current at 24 VDC ......................................... 1 A
Marine approval:
Det Norske Veritas, Ships & Offshore ............. Standard for Certification No. 2.4
Observed authority requirements: Standard:
EMC 2004/108/EC .................................................. EN 61326-1
LVD 2006/95/EC ...................................................... EN 61010-1
UL, Standard for Safety ......................................... UL 508
EAC TR-CU 020/2011 ............................................ EN 61326-1