Tube, Accu, & Calibration
Choose
Tube
you want to calibrate in the ’Tube’ menu (A-J)
Up to 10 calibrations can be stored
Before each infusion, be sure the correct
Tube
appears in
the top left corner so that the correct infusion settings are
used
When calibrating we strongly recommend using a gram
scale that measure in 0.1g increments or a graduated
cylinder when fine tuning motor speed. If you do not have
either of those tools, infuse into a container and draw up
fluid infused with a syringe, or connect a female/female
transfer device to the male end of your admin set and a
syringe.
The ‘Accu’ Screen