Practivet INFUSION PUMP - Calibration Process; Fine Tune Based On Rate; Accum Volume Coefficient Adjustment

Practivet INFUSION PUMP
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Process
1. Set up a way to measure volume infused in 0.1
mL increments (empty container & draw up
fluid with syringe, gram scale, etc.)
2. In the settings interface, go to ‘ACCU
3. Press OPTION 5
4. Prime administration set and feed primed line
through pump
5. Put male end in measuring container
6. Press START
7. Wait until calibration is finished (Vol 10mL)
8. Input amount infused next to ‘Real’ (This value
should be between 5 & 7)
9. Press ENTER to save this value
Option 1, 2, 3, 4: ml/h Adj
Fine Tune Based On Rate
1. Run a 10ml infusion at each rate: 25, 150, 600,
& 1100 ml/h
2. Input percent accuracy for each
If output was 9.2 ml you would input -8%, if it
was 10.3 ml, you would input 3%
Keep in mind this is OPPOSITE of what you
would think!
Option 6: Accum Vol Coef Adj
Please only make adjustments to Option 6 at the
direction of a PractiVet Product Specialist