Precision Medical Live Active Five Replacement Instructions

Precision Medical Live Active Five
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Sieve Bed
Assembly
Replacement
Instructions
300 Held Drive
Northampton, PA 18067
precisionmedical.com
1-800-272-7285
Turn concentrator o, and
remove the battery and external
power from unit.
Turn unit upside down, and place
on soft surface. Remove screws
and bottom cover
Gently push until sieve bed
replacement is fully seated, flush
with bottom surface. Replace bottom
cover and secure screws.
Turn unit upright and insert
battery. Gently push until battery
is locked in place.
Register the Sieve Bed Assembly in the Concentrator
Press and hold the Audio Pause button to enter the menu. Press the 5 button to navigate to the Sieve Beds
option. Press the 3 button to select the Sieve Beds option. Press the 3 button again to register the sieve bed
assembly. “Done” will appear on the screen once completed. Press the Audio Pause button to exit the menu.
Pull upward on the Pull String Loop
while holding the device down to
remove sieve bed assembly.
Remove 4 plugs from new sieve
bed assembly, and slide it into the
concentrator.
1
4 5
2 3
6

Other manuals for Precision Medical Live Active Five

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Precision Medical Live Active Five and is the answer not in the manual?

Precision Medical Live Active Five Specifications

General IconGeneral
BrandPrecision Medical
ModelLive Active Five
CategoryMedical Equipment
LanguageEnglish

Related product manuals