1. TECHNICAL DESCRIPTION
The Lumina Max light is a medical device, class I, in compliance with MDD (Medical Devices Directive)
section IX. The light must only be used in dental clinics by dental personnel, and only for designated
purposes. The product is classied as an electronic medical device electric class II according to DS/EN
60601-1:2006. The product is classied as a class 2 LED product according to EN 60825-1:2001 Ed1.2.
Dentalica declares that the product described in this manual complies with the above mentioned
standards. The Lumina Max light consists of a hand piece, a docking station and an external power
supply. Disposable covers and tips protect against cross contamination.
Caution
1. Do not expose directly into the eyes, or look directly at the
emitted light.
2. Eyewear protection should be used.
3. Medical electronic equipment is sensitive to electromagnetic
elds. Portable and mobile high frequency communication devices
may interfere with medical devices. In order to maintain proper
operation, follow instructions in this manual.
4. Use of the light without a tip may cause overheating of the tissue if the light is used close to the
gingiva.
5. Extended use cause tip and cover to be hot and may cause slight burning sensation of mucosa, if
touched.
6. Do not reuse covers or tips as this may cause cross contamination.
7. If the light beeps every 2s when placed in docking station, it is a warning that the charging voltage is 6
V or higher, i.e. a wrong power supply is being used.
POWER SUPPLY
Input power 100240VAC, 50/60Hz
Output power 5.0VDC, 1.0A
BATTERY
Battery type LiFePO
4
Nominal voltage 3.2V
Nominal capacity 1.200mA
Re-charging time 2 hours
User capacity approx. 500s
LED DIODE
Output intensity 5.000-6.000 mW/cm
2
Wavelength 450-470nm (85%),
picco 460nm
OTHER
Operation mode Intermittent.
No continous use over 1
minute to prevent over
heating