Pressure Systems, Inc. Model 9046 User’s Manual
Page 38 www.PressureSystems.com
7 single binary float to single binary float (big endian: msb first) 4
8 single binary float to single binary float (little endian: lsb first) 4
Example:
Command module to return decimal raw "Temperature/Resistance" A/D counts data for Primary
Transducer channels 1, 5, 9, and 13:
"
a11110"
Response contains data for channels 13, 9, 5, and 1 (left to right):
"
32767.000000 -32700.000000 10.000000 16385.000000"
Please note that channel 13 is saturated at +full scale and channel 9 is almost saturated at -full
scale. Channel 5 reads near zero and channel 1 is about 1/2 +full-scale.
Note
NUSS does not use this command normally. You may execute it
manually via the ’Execute Commands’ or ’Execute Script’ functions
of NUSS - though this should be avoided if NUSS is currently
operating the module via other NUSS functions. NUSS normally
reads all the data it needs from a module by configuring
autonomous streams (see the ’c’ command).