(105) Sulphite (HR)
0 - 300 mg/l (Na SO )
2 3
Sulphite HR N° 1 (TbsHSULFHR1…)
Sulphite HR N° 2 (TbsPSULFHR2…)
10 ml
24 mm
4
4
4
1) Fill 10ml of the sample water into a clean 24mm cuvette.
2) Replace the screw lid on the cuvette and insert it into the PrimeLab
sampling chamber. Align the arrow on the front of the cuvette. Press
ZERO.
3) Remove the cuvette from the chamber and unscrew the lid.
4) Add a "Sulphite HR N° 1" tablet to the water sample in the cuvette.
5) Crush the tablet with a clean stirrer until it is completely dissolved.
6) Add a "Sulphite HR N° 2" tablet to the water sample in the cuvette.
7) Crush the tablet with a clean stirrer until it is completely dissolved.
8) Replace the lid on the cuvette, swivel it carefully and insert it in the
sampling chamber of the PrimeLab. Align the arrow on the front of the
cuvette. Press TEST.
9) After the lapse of a 2 minute countdown the determined result is
displayed. Press button 4 to save the test result together with date and
time on the device.
10) Press button 3 to convert this into various measurement units (*).
Name on device: 105-Sulfite-HR-tab
Tablet
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO = continue with step 2)
-> Use last ZERO = continue with step 4)
1 2 3 4
Continued...
!
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.