245
245
Notes
-> (*) Conversion of the measurement result to SO
3
-> Filter sample if neccessary to obtain a clear solution.
-> Cell, lid and stirring rod need to be cleaned immediately after to prevent
staining.
-> Expect low results if tannin or tannic acid is present.
-> Chlorine >250 mg/l
Nitrite >200 mg/l
Iron >20 mg/l
Sulphide >10 mg/l
cause interferences.
Continued...
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
(105) Sulphite (HR)
Name on device: 105-Sulfite-HR-tab
Tablet
0 - 300 mg/l (Na SO )
2 3