999016 – Revision G P a g e | 34 of 44
11.0 General Inspection, Maintenance and Cleaning
11.1 Service
No service is to be carried out on the Hoist while transferring a person to reduce the risk of injury.
Service must be completed by a Prism Medical UK an approved trained professional.
Do not attempt to service the product yourself, or warranty is void.
To ensure the safety and continued good function of your Hoist, routine service must be performed on your Prism CP Hoist.
Service should be completed by a Prism Medical UK approved trained professional every 6 months to ensure the products
required standard is maintained. The service history of the product should be documented each service in the Service Log at the
back of this User Manual.
When the Hoist is serviced, the 6 month service checklist must be completed for the Prism CP Hoist.
Service Manual Document Number: 995016.
Spare Parts Manual Document Number: 992016.
The Service must be completed every 6 months after installation of the Hoist to comply with LOLER Regulations.
The Prism CP Hoist has an expected Service Life of 10 Years.
Contact your local authorised Prism Medical UK dealer if you:
• Need more information.
• Have any questions about the use or service of your Hoist.
• Notice any change in the performance.
• Want to report an unexpected occurrence.
• Want to arrange a service.
• Need to ascertain necessary information for replacement parts and components.
Contact details of your local Prism Medical UK dealer are shown on the last page of this manual.
11.2 Inspection
Inspection is to be completed prior to each use by the user of the Hoist.
Should any of the components in the table below fail the inspection, DO NOT use the Hoist.
Contact your local authorized dealer for service – contact details are on the last page of this manual.
Ensure all component inspections in the Table below are completed prior to each use of the Hoist.
Check List before Use