8
discontinued and a corresponding error code is displayed.
Error Code Illustration Possible cause(s)
Err 2
Can’t detect the pulse.
Err 4
The ination time is too long, the cu
is not worn properly or the tube has
been disconnected.
Err 5
The measurement result is invalid.
Please measure again.
Err 6
There is excessive interference.
Err 7
Cus inner pressure is > 260 mmHg.
Err E
Please return the monitor for repairing.
Hi
When pressure is ≥ 260 mmHg, “Hi” is
displayed.
6.2. Troubleshooting
Problem Solution
The display remains blank when the
device is switched on, although the
batteries are in place.
Check batteries for correct polarity
and if necessary insert correctly.
If the display is unclear, reinsert
batteries or replace them.
The device frequently fails to measure
the blood pressure values, or the values
measured are questionable.
Check the positioning and t of
the cu.
Measure the blood pressure again
in a quiet environment , following
the steps outlined in section 3.
Blood pressure measured diers
from those values measured by the
doctor. Every measurement produces
a dierent value although the device
functions normally and the values
displayed are normal.
Attempt to measure your blood
pressure at the same time of day, as
blood pressure changes during the
course of the day.
Record the daily values and consult
your healthcare professional.