EasyManua.ls Logo

Pro PRO900 - Calibration and Maintenance; Calibration Procedure

Pro PRO900
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
ver. 1.1
5. Calibration
All devices are factory calibrated. However, due to mechanical shock
or exposure to too high or too low temperatures, the need to recalibrate
the device may occur.
PRO900 Digital Spirit Level calibration
a) Find a horizontal surface. You can check the surface using
the built-in spirit level. If the digital level is not equipped
with a spirit level, use another spirit level for measurement.
b) Press the „HOLD” (5) button for approx. 3 seconds.
The “CAL1” sign will appear on the screen.
Press the „REF” (8) button.
c) Leave the digital level intact until the screen
shows a „CAL2” indication.
d) Rotate the digital level 180° and place the device
on the very same horizontal surface.
Press the „REF” (8) button.
WARNING! Calibration can be repeated several times. While calibrating
the device, it is important to place the device ideally
in repeatable conditions in horizontal position.
Calibrating the device on an inclined surface may cause
wrong indications.
e) Wait approx. 4 seconds until the screen shows:
Calibration has been finished correctly!
180°180°