5.3. Basic Measurement – Pulse Velocity
In order to measure the pulse velocity it is necessary to measure the path length between the
two transducers.
Select
“Basic Measurements”
Select parameter setting “path
length”
Enter the path length
Start the measurement
The display shows:
- Transmission time
- Measured pulse velocity
- Received signal level *
Stop the measurement
(Not necessary in burst mode)
Save the result or:
Adjust the gain settings.
(See Chapter 3) or:
Start a new measurement
* Best results are obtained when the received signal level is around 75%. Use the automatic
gain and voltage settings (See Chapter 3) for optimized performance.