Step 2: Inhalation
Hold the nebulizer in your hand stably and start
inhalation with relaxed posture by placing lips on or
very close to the mouthpiece. See Figure 17. Do not
cover the holes of the mouthpiece when you inhale;
otherwise, the nebulization rate will decrease. See
figure 18.
ATTENTION:
• If the device does not detect any medication in
the medication chamber, the power indicator
turns red and it will shut off automatically.
• During treatment, you may adjust the nebulizer
to any angle in order to be more relaxed.
However, make sure the medication stays in
contact with the mesh; otherwise, the nebulizer
will shut off automatically after a few seconds.
• When the medication is close to depletion,
it is recommended that you tilt the nebulizer
slightly towards you. This allows the remaining
medication to contact the mesh to nebulize.
• Provide close supervision when the nebulizer is
used by children or compromised individuals.