Page #28 background imageLoading...
Page #28 background image
26
27
Prodigy Diabetes Care, LLC 1.800.243.2636
PRODIGY Mini-Mist
Owner’s Manual
System TroubleshootingSystem Troubleshooting
v The unit is on, however, it nebulizes
weakly or is taking too long for treatment
• Make sure the medication chamber is
properly installed. See page 10.
• Follow directions for cleaning after each use.
See page 20.
• Follow directions for disinfecting. See page 19.
• Replace batteries per installation instructions.
See page 13.
• Nebulization rate may vary based on the
medication used.
v The nebulizer shuts off during usage
• Adjust the nebulizer’s angle so the medication
is in contact with the mesh.
• The medication has run out and the
medicationchamberneedstoberelled.
See page 14.
• The medication chamber is loose and
not completely attached. Re-attach the
medication chamber correctly and restart the
power. See page 10.
• Batteries need to be replaced. See page 13.
v The nebulizer does not shut off automatically
• Electrodes on the medication chamber are
clogged with medication or water and need to
be cleaned. See page 22.
• Electrodes on the nebulizer and medication
chamber are stained. Remove the stains and
restart the power. See page 22.
If your nebulizer still does not function properly,
please contact Customer Care at 1.800.243.2636
for 24-hour assistance.