25
Rev. C / January 2012
PRS-812 Resistance Meter
3. An LED corresponding to the measurement’s order of magnitude will be energized
Measuring 958Ω in Auto
Ohms Mode
Measuring 958Ω in Auto
Exponenal Mode
F. Measurements from 1.0x10
4
to less than 1.0x10
6
ohms:
1. When the resistance measurement is greater than 1.0x10
4
ohms test voltage will automacally
be increased to 10 volts, and the resistance range will be adjusted as necessary.
2. A stable resistance measurement will be obtained within 2.0 to 4.0 seconds.
3. Once a stable measurement is conrmed and displayed, will be energized in the LCD and
the electricaon mer will stop.
Ω
-
Electricaon Remains 2.0 – 3.0 sec.
on Measurements <1.0E+06 Ohms