Sound Level Meter User Manual
15-Calibration potentiometer(CALL)
For external standard level calibration adjustments.
16-Tripod mounting screw
17-Battery cover
18-Microphone
1/2 inch Electret Condenser microphone
5-Calibration Procedures
Make the following switch settings:
Frequency weighting: A-weighting
Time weighting: FAST
Level range: 50~100dB
Insert the microphone housing carefully into the 1/2 inch insertion hole of the calibrator(94dB@1kHZ)
Turn on the switch of calibrator and adjust the CALL potentiometer of the unit 94.0dB is displayed.
NOTE: All products are well calibrated before shipment. Recommended recalibration cycle: 1 year.