Page 4 of 34
QSA Global, Inc. 40 North Avenue Burlington, MA 01803 888.272.2242 781.272.2000 F: 781.359.9179 qsa-global.com
EQUIPMENT SPECIFICATIONS.
Table 1 OpenVision™ DX Specifications
Imaging Area (Field of View) 4 in. x 6 in. (10 cm x 15 cm)
Dimensions
X-Ray Tube to Imager
Throat Depth
Imager Thickness
X-Ra
Tube Thickness
9.5 in. (24 cm) to 20.8 in. (53 cm)
21 in. (58 cm)
2.5 in. (6.4 cm)
3.4 in.
8.7 cm
Startup Time 30 seconds
Shutdown Time 5 seconds
X-Ray Energies
Low
Medium
High
Voltage (kV)
40
55
70
Current (mA)
0.300
0.218
0.171
Dose @ 12 in (30.5 cm) from beam port
50 R/hr (0.5 Sv/hr)
80 R/hr (0.8 Sv/hr)
95 R/hr (0.95 Sv/hr)
Beam Collimation (Approximate)
18° Horizontal
10° Vertical
Image Resolution 250 microns (0.01in)
Battery Life
Continuous duty with 5 Ah battery
Standb
3 hours
8 hours
System Weight
C-arm
Monitor w/ handle (optional)
Packa
ed s
stem
shippin
wei
ht
15 lbs. (6.8 kg)
2 lbs. (.9 kg)
50 lbs.
23 k
Operating Temperature
-20° F to 120° F
-29° C to 49° C
Storage Temperature
-20° F to 140° F
-29° C to 60° C
Display Options
HDMI Monitor: 7 in. LCD, 1920 x 1200 pixels
WiFi Tablet: 10 in Touchscreen, 1400 x 900
PirateEye: 854 x 480 pixels
lso compatible with an
device that accepts an HDMI si
nals
Recording (OpenVision DX)
Internal storage
Resolution
Ima
e / video transfer
128 GB
1280 X 720
USB
Recording (WiFi Tablet)
Internal storage
Resolution
Ima
e / video transfer
4 GB with expandable 128 GB memory
1280 X 720
USB
Shipping Dimensions
32 in. x 20.5 in. x 12.5 in
82 cm x 52 cm x 32 cm
FDA Accession # 1680071-000