31. After solving equation (13) for V
R
micro use this value in equation (14) to calculate V
c
micro
for the micro sample cell.
(
)
(
)
- micro
P
micro/
P
micro
V
+ micro
V
= micro
V
21
RcalC
1
14)
NOTE: If only the large sample cell is to be used, it is not necessary to recalibrate the small and
micro cells. Similarly, if only the small cell is to be used, it is not necessary to calibrate the micro
cell.
If the sample is analyzed at a temperature different by more than 4°C from that used to calibrate the
instrument, the MULTIPYCNOMETER should be recalibrated at the new temperature.
Calibration Spheres
The table below shows the various spheres that are available for calibrating the instrument.
PART NUMBER
SIZE DIAMETER (mm) VOLUME (cm
3
)
01500-MICRO
1
Micro
2
12.7 1.0772
01500-SMALL
Small 23.812 7.0699
01500-MEDIUM
Medium 38.1 28.958
01500-LARGE
Large
2
47.625 56.5592
NOTE: The large and small calibration spheres are available with NIST measured diameters.
Contact qc.sales@quantachrome.com for more details.
____________________________________________________________________________
1. Sold in pairs.
2. Provided with the Multipycnometer.