Chapter 5 Service Instructions
Revision W HF Series X-ray Generators - Service Manual
5-16 Quantum Medical Imaging, LLC
Error -
Could not
open file
{text
string}
There was a problem loading the
selected APR tree. Note: {text
string} will contain additional
details.
1. Wrong file was selected.
2. File is corrupt.
Error -
Only one
instance of
the appli-
cation can
be run at a
time.
Opening up of a second instance
of the same program was
attempted by the user.
Only one instance of the application
can be run at a time.
Expose
WDT
Exposure sequence disabled due
to loss of communication with
EXPOSE watchdog timer.
Call Technical Service/Support.
Exposure
Lock
The CR/DR system is blocking the
exposure for:
1. Trying to expose a body part
that has already been exposed.
2. The CR/DR system is not in an
acquisition mode.
1. Place the system in valid acquisi-
tion mode.
2. In case of Emergency, press
URGENT button to override expo-
sure lock.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action