5. Press and hold the pulse switch to trigger a sequence
of therapeutic light pulses� This sequence takes a few
seconds, depending on the energy level setting (see
specifications)
If during the pulse sequence the Light Unit Assembly
feels too hot on the skin, remove it immediately and wait
until the ready indicator lights up again�
6. After emitting a pulse sequence, remove the handpiece from the skin for about 10
seconds to allow the Light Unit Assembly to cool down� Apply the next sequence
only when the Ready indicator lights up�
Placing the handpiece on the skin before the Ready indicator lights up may result
in unwanted side effects due to pre-heating of the skin�
7. Sequentially administer pulses on the treatment area
from right to left, in a horizontal pattern, with no overlap
Stacking and/or overlapping may cause crusting or
blisters which could result in transient pigmentation
changes�